Keterangan
CyberScan PCD 6500
Product Features
Top-of-the-line, research-grade benchmeter
Measures up to 10 parameters
Up to four channels measured and displayed simultaneously:
- Two channels of pH/ mV/ ISE inputs
- One channel for pH ( FET)
- One for Dissolved Oxygen/ BOD/ OUR/ SOUR
- One channel for Conductivity/ Resistivity/ TDS/ Salinity input ( 2-cell or 4-cell)
Advanced design ensures adequate shielding to prevent cross talk or other interference.
Wide pH and mV measurement range:
-2.000 to 20.000 pH and ± 2000.0 mV
‘ Direct/ Indirect’ , ‘ Known Addition/ Subtraction’ and ‘ Analate Addition/ Subtraction’ Ion measurement methods in ppm, % , mg/ mL and mole/ L.
DO measurement includes five-day BOD testing, OUR, and SOUR. Simple DO measurement ranges from 0 to 60 mg/ L and 0 to 600% saturation.
Single point and Multi-point standardisation for Conductivity, Resistivity, TDS and Salinity measurements using two-cell and four-cell probes. Built-in replatinising circuit.
Advanced real-time on-screen graphing function provides useful indication for specific measurements such as titration.
Extensive setup screens enable you to customise meter to your needs, e.g. Isopotential Point, Temperature Units, Ion Method, ISE Electrode Type, BOD/ OUR/ SOUR Configuration, Cell Constant, Temperature Coefficient, TDS Factor, Alarm Limits and other functions.
Windows® CE-driven, full-color touchscreen provides unmatched ease of use in operations and setups with user-friendly icons, user prompts and context specific ‘ help’ screens.
Advanced USB, IRDA connectivity allows extensive host/ device communication capabilities.
Ethernet/ Internet-ready with RJ45 input for connection to LAN or Internet for quick online research and data sharing.
Up to ten thousand data set storage per user login ( currently one of the highest in the industry)
Password protected access and customisation for up
Applications
Laboratory/ Industrial: Use for water quality testing, sewage treatment, food processing ( for HACCP compliance) , geological testing. Great for R& D and QC applications in the food and pharmaceutical industries, paper/ pulp processing, chemical industries and related fields. Suitable for most ASTM and EPA test procedures.
Educational: Perfect for research labs, suitable for all types of chemistry labs and ecology studies.
Specifications
pH Range -2.000 to 20.000 pH
Resolution & Accuracy 0.1/ 0.01/ 0.001 pH & ± 0.002 pH
mV Range -1800 to + 1800 mV
Resolution & Accuracy 0.1 mV & ± 0.1 mV
Ion Concentration Range 1 x 10-7 to 9.99 x 1010
Resolution & Accuracy 4 sig. figures & ± 0.17n%
DO Range 0 to 60 ppm, 0 to 600%
Resolution & Accuracy 0.01/ 0.10% & 0.1% + 1 LSD
Conductivity Range 0 to 500 mS/ cm
Resistivity Range 3 ohm cm to 100 megohm cm
Salinity Range 0 to 90 ppt
Accuracy 0.5% full scale + 1 LSD
Cell Constant 0.1, 1.0, 10.0
Temperature Range -5.0 to 105.0 ° C ( Meter)
Resolution & Accuracy 0.1 ° C & ± 0.2 ° C
No. of pH Calibration Points Up to 5
pH Buffer Options USA: 2, 4, 7, 10, 12
NIST: 1.68, 4.01. 6.86, 9.18, 12.45
Euro: 1, 3, 6, 8, 10, 12Â
No. of Ion Calibration Points 5
Incremental methods KA, KS, AA, AS
Temperature Compensation Automatic / Manual ( 0 to 100 ° C)
DO Testing 5-day BOD, OUR and SOUR testing
Automatic Barometric Pressure Compensation 450 to 825 mmHg
Memory 10, 000 Data Sets with Date/ Time stamping
Programmable Data Storage/ Output On stable, time and manual
Output USB, IRDA, RS232C
Print Interval 3 to 86400 sec
Electrode Input FET ( solid-state non-glass) or Glass pH electrode
Display 640x480 digit LCD
Power 9VDC, 3.3A center negative
Price : USD 4, 372.00
Order No.
ECPCD650044S
Color Touch Screen CyberScan PCD 6500 - 4 channel pH/ pH FET/ mv/ Ion/ Conductivity / Resistivity/ TDS/ Salinity/ DO/ BOD/ OUR/ SOUR RS232, USB, IRDA Meter with pH electrode EC620130, Conductivity electrode EC620165, ATC probe, DO probe, electrode holder & universal power adapter ( US, UK, EURO, Japan )
ECPCD650042S
Color Touch Screen CyberScan PCD 6500 - 4 channel pH/ pH FET/ mv/ Ion/ Conductivity / Resistivity/ TDS/ Salinity/ DO/ BOD/ OUR/ SOUR RS232, USB, IRDA Meter with DO probe, electrode holder & universal power adapter ( US, UK, EURO, Japan )
ECDAS6000